Zero and Span Calibration: What Actually Happens, Step by Step
A gas detector can carry a fresh calibration label and still read low at the top of a ballast tank. Usually nothing’s broken. It was zeroed in air that wasn’t clean, spanned against a cylinder that didn’t match the value in its menu, or fed gas at the wrong flow, and each of those still ends with “pass” on the screen. Here’s what zero and span actually set, and what to check at each step.
What zero and span actually set
A sensor doesn’t produce a gas reading. It produces a small electrical signal, which the instrument converts into ppm, %LEL or %vol using a line stored in memory. Zero fixes where that line starts and span fixes how steeply it rises, which is why portable detectors normally get by with a two-point calibration.
MSA’s documentation puts it simply: zero calibration resets the baseline reading to zero, and span gas of a known concentration adjusts the reading to that known value. Because the span is measured from the zero, the order matters. If the zero is off, the slope worked out during the span is off too, and the error carries into every reading afterwards. Honeywell’s calibration guidance says the same: zero first, then span.
There’s no IMO procedure for any of this. SOLAS regulation XI-1/7 requires a suitable means of calibration and leaves the method to the manufacturer, so the manual is the procedure and the details below vary by model. Our article on what MSC.581(110) means for your detectors covers the regulatory side.
Bump test, calibration check and full calibration
US OSHA guidance describes a bump test as a qualitative check that gas reaches the sensors and the alarms work, and it says plainly that a bump test doesn’t measure accuracy. A calibration check applies a known, traceable concentration and confirms the reading falls within the manufacturer’s tolerance, without adjusting anything. A full calibration is the zero and span adjustment this article is about.
So a detector that passes its daily bump can still read some way off the cylinder value. Our guide to bump test vs full calibration and what PSC inspectors check covers the difference in full.
Before you start: gas, regulator and warm-up
Most calibration errors are locked in before the first button is pressed. Start with the gas: the mix and concentration the manual specifies, certified, traceable and in date. A typical 4-gas detector takes one quad-mix cylinder, such as a 4-gas calibration mix of CO 50 ppm, H₂S 25 ppm, 2.5% methane and 12% O₂. Extra sensors may need a second cylinder from the calibration gas mixtures range.
The regulator depends on how the detector samples. Diffusion detectors take gas through a calibration cap from a fixed-flow regulator at the manufacturer’s rate; MSA designs around 0.25 L/min while 0.5 L/min is common elsewhere, so a regulator borrowed from another kit isn’t always right. Pumped detectors need a demand-flow regulator, which releases gas only as the pump draws it. A fixed-flow regulator on a pumped unit can deliver too little or too much gas.
Then give the detector time to stabilise in clean air, for as long as the manual says. Gas applied to a sensor that’s still settling calibrates it against a moving baseline.
Step 1: take as-found readings
Before anything is adjusted, apply the span gas and write down what each sensor reads. That’s the as-found reading, and it’s the only record of how far the detector drifted since its last calibration. Once the span adjusts the sensor, that evidence is gone.
Automatic routines often go straight to adjustment, so this usually means running a calibration check first. A large as-found error is what tells you the interval may be too long or a sensor is ageing; see why sensor readings drift over time.
Step 2: zero
Zeroing tells each sensor that what it’s sampling contains none of its target gas, and the instrument takes your word for it. Industrial Scientific points out that if the air isn’t truly clean, any relevant gas in it gets stored as zero, and readings stay wrong until the unit is zeroed again in fresh air or with zero air.
On a ship that rules out the engine room, the pump room and anywhere near an exhaust; open deck on the upwind side usually works. Where clean air can’t be confirmed, Honeywell’s guidance is to zero with zero-grade air or high-purity nitrogen instead. Nitrogen reads 0% oxygen, though, so the manual decides which channels it’s used for; many instruments set the O₂ sensor to its fresh-air value (20.9%, or 20.8% on some MSA models) rather than to zero.
CO₂ is the main exception: outdoor air holds roughly 425 ppm, so some makers won’t zero a CO₂ channel in fresh air at all. See how CO₂ channels are zeroed and spanned.
On an MSA Altair 4X the zero takes about ten seconds and ends with ZERO PASS, or ZERO ERR with the failed sensor flagged. Sort out a zero error before spanning.
Step 3: span
With the cap fitted and gas flowing at the right rate, wait for the reading to settle (about a minute per sensor in MSA’s table, two minutes or until stable for other makers), then confirm. The instrument recalculates the slope so that reading matches the span value.
That span value is where calibrations quietly go wrong. MSA’s manual says the sensor span value in the instrument should match the concentration on the cylinder, unless an LEL simulant gas is being used. Suppose the menu still says 25 ppm H₂S from the last cylinder but the new one holds 20 ppm. The detector is set to read 25 when it sees 20, so it reads about 25% high from then on.
The LEL channel adds two complications. First, simulant gas: catalytic sensors scaled for pentane are usually spanned with methane, and MSA’s calibration table spans a pentane-scaled sensor with 1.45% methane at a span value of 58% LEL. A certificate showing 58% LEL against a methane cylinder isn’t a mistake.
Second, the LEL basis. The US definition (ISO 10156) treats 5.0% vol methane as 100% LEL, while IEC and EN regions use 4.4%, so the same cylinder is a different %LEL depending on the detector’s setting:
| Methane cylinder (% vol) | %LEL on the 5.0% basis | %LEL on the 4.4% basis |
|---|---|---|
| 1.45% | 29% | 33% |
| 2.2% | 44% | 50% |
| 2.5% | 50% | 57% |
Cylinders labelled “2.5% methane, 50% LEL” are on the 5.0% basis; our own 2.5% methane calibration cylinder gives methane’s LEL as 5.0% on its page. If the detector is configured for 4.4%, MSA’s table gives a span value of 57% for that cylinder. Enter 50% instead and the detector reads roughly 12% low on its own scale. Neither basis is wrong, but the cylinder, span value and detector setting have to agree.
And where methane might be present, calibrate with methane, since a degrading sensor calibrated on propane or pentane can under-read it.
Step 4: verify, check alarms and record
Turn off the gas, remove the cap and let the sensors return to baseline in fresh air. One that won’t come back close to zero needs looking at before the detector goes back into service.
Then check alarm set points and the audible, visual and vibrating alarms, plus pump flow and the actual sample hose on a pumped unit, since a leaking or blocked line changes what reaches the sensor. Record the as-left reading for each sensor.
The certificate should show the instrument and each sensor, the gas concentrations with cylinder lot and expiry, as-found and as-left readings, alarm results, the next due date and who did the work. Adding the span value and LEL basis makes methane figures easier to check later. See what PSC inspectors look for on calibration certificates.
When a span won’t take
Rule out the setup first: an expired or wrong cylinder, a span value that doesn’t match the cylinder, the wrong regulator for the sampling method, a cap that isn’t seated, or a blocked filter. If all of those check out and the sensor still fails, or only passes by using up most of its remaining adjustment, it’s probably near the end of its life. Our article on when to replace a sensor instead of recalibrating covers how to tell.
The sequence in short
The manual governs the detail for each model, but a full calibration generally runs like this:
- Confirm the gas is the right mix, in date, and matches the span values set in the instrument.
- Fit a fixed-flow regulator for a diffusion detector or a demand-flow regulator for a pumped one.
- Switch on and let the detector stabilise in clean air.
- Apply span gas and record the as-found reading for each sensor.
- Zero in clean air, zero air or nitrogen, as the manual specifies for each channel.
- Span each sensor until the reading is stable, checking the LEL basis and span value.
- Return the detector to fresh air and confirm it comes back to baseline.
- Check alarms, pump flow and the sample line.
- Record as-found and as-left readings, gas details and the next due date.
How often to repeat it depends on the manufacturer and your SMS; our calibration interval table gathers the common references.
Getting it done on board in Singapore
For onboard gas detector calibration in Singapore, our engineers board alongside or at anchorage with the gases your sensors need. They record as-found readings before any adjustment, zero in clean air or nitrogen for O₂-free checks, span against traceable gas, and test alarms, pump flow and sample lines before the certificate is issued. To arrange a visit, send us the vessel name, IMO number, berth or anchorage, ETA and a list of your detectors with make, model and serial numbers. If you calibrate in-house, span gas and regulators are supplied from Singapore too; see our ship fixed and portable gas detection services.