Skip to main content

Alrightes

Machinery & cargo instruments · Singapore

Pressure & Temperature Instrument Calibration in Singapore

On-board calibration of pressure gauges, switches, transmitters and temperature sensors on ships alongside or at anchorage in Singapore. We test against traceable reference standards, check the alarm and shutdown set points they feed, and issue a certificate with as-found and as-left readings.

  • ISO 9001:2015
  • ISO 45001:2018
  • bizSAFE Level 4
  • bizSAFE Star
  • Alongside & anchorage
  • As-found / as-left certificate
Marine engineer calibrating a pressure gauge on board a ship in Singapore
Gauges · switches · transmittersPlus RTDs, thermocouples, thermometers and the alarms they trigger

Pressure and temperature calibration in Singapore, in short: Alright Engineering Solutions (AES) calibrates pressure gauges, pressure switches, pressure transmitters, thermometers, RTDs and thermocouples on ships in Singapore. Each instrument is compared with a traceable reference at several points, adjusted where it can be, and recorded as found and as left. Where an instrument feeds an alarm or shutdown, we check that the set point trips where it should.

Scope

Pressure and Temperature Instruments We Calibrate

From the engine room gauge board to the cargo control room, one attendance can cover the instruments your planned maintenance system says are due.

Pressure gauges

Bourdon tube and digital gauges on main and auxiliary engines, boilers, compressors, pumps and hydraulic systems, plus test gauges.

Pressure switches

Low lube oil, low cooling water and high pressure switches, checked for switching and reset points against the alarm list.

Pressure transmitters

4–20 mA and smart transmitters checked at the sensor and at the alarm monitoring system display, so the loop agrees end to end.

Temperature sensors

PT100 RTDs and thermocouples, including exhaust gas, jacket cooling water, lube oil and cargo pump temperature monitoring.

Indicators & controllers

Temperature indicators, transmitters and controllers, plus thermometers and temperature gauges on local panels.

Calibrators & test tools

Pressure calibrators, manometers, temperature calibrators, multimeters, torque wrenches and tachometers carried on board.

Method

How Each Instrument Type Is Checked

A single reading at one point tells you very little. We check across the range the instrument actually works in. More detail in our guide to pressure gauge, switch and transmitter calibration for vessels.

InstrumentReference usedWhat we check
Pressure gaugeCalibrated reference gauge or pressure calibrator on a hand pump or comparatorReadings at several points rising and falling, to catch error and hysteresis
Pressure switchPressure calibratorPressure at which the contact changes over and resets, against the set point
Pressure transmitterPressure calibrator plus mA meterOutput at 0, 25, 50, 75 and 100% of range, and the AMS reading
RTD / PT100Reference probe in a dry-block or bathResistance or reading against the reference at several temperatures
ThermocoupleReference probe or temperature calibratorSensor response, compensation and indicator reading
Thermometer / temp gaugeReference probe in a dry-block or bathIndication at several points across its scale
Alarms & shutdowns

Most of These Instruments Protect an Engine, Not Just Show a Number

A low lube oil pressure switch that trips late, or an exhaust temperature sensor reading 20°C low, won't show up on a walk-round. It shows up when the engine is already in trouble. That's why we calibrate the sensor and then confirm the alarm or slowdown it feeds.

  • Switch and transmitter set points checked against your alarm and shutdown list
  • Sensor reading compared with what the AMS or local panel actually displays
  • Faulty sensors, capillaries and cabling flagged with the readings to back it up
  • Records ready for the machinery and alarm system checks at class surveys

Not sure what's due? Our marine instrument calibration interval table sets out common intervals by instrument type.

How it works

How an On-Board Calibration Job Runs

Planned around your port stay and the machinery that can be isolated.

  1. Send the list

    Tag numbers, ranges and locations from your PMS, plus ETA, ETD and berth or anchorage.

  2. Plan isolation

    We agree with the chief engineer which systems can be isolated and when, so nothing critical is taken out at the wrong time.

  3. As-found test

    Each instrument is checked against the reference before any adjustment, so the drift is on record.

  4. Adjust or replace

    Zero and span are adjusted where the design allows. Instruments outside tolerance that can't be adjusted are flagged for replacement.

  5. Loop & alarm check

    Transmitters are checked through to the AMS, and switches through to the alarm or shutdown they trigger.

  6. Certificate & labels

    As-left readings go on the certificate, each instrument gets a calibration label, and the list is ready for your PMS.

Large gauge list? Send it ahead and we'll plan the attendance around your port window.

On board or in the workshop

On-Board Calibration vs Sending Instruments Ashore

Fixed instruments are best done where they're installed. Portable test tools can go either way.

On boardAshore
Best forFixed gauges, switches, transmitters and sensorsPortable calibrators, test gauges and spare instruments
Loop and alarm checkYes, through to the AMS and shutdownsNo, instrument only
Removal neededUsually not, or only brieflyYes
Time on boardPlanned around isolation windowsNone for the ship
Common problems

What We Find When Ship Instruments Are Tested

Gauges reading high at zero

Vibration and pressure pulses bend the Bourdon tube and shift the pointer. The gauge looks fine but reads wrong.

Switches tripping late

Worn diaphragms and stiff contacts move the switching point away from the alarm set point.

Transmitter vs AMS mismatch

The transmitter is correct, but scaling in the AMS isn't, so the display and the real pressure disagree.

Thermocouple drift

Exhaust gas thermocouples age in high heat and start reading low, hiding a cylinder running hot.

Damaged capillaries

Kinked or broken capillary lines on temperature gauges give slow or frozen readings.

Blocked impulse lines

Sludge in gauge lines and snubbers dampens the reading or stops it moving.

Documentation

What's on Your Calibration Certificate

  • Instrument tag, make, model, serial number and range
  • Reference standard used, with its own certificate details
  • As-found and as-left readings at each test point
  • Switch set and reset points, where applicable
  • Pass or fail against the stated tolerance
  • Date, vessel, next due date and engineer's signature

See what inspectors check in our guide to marine calibration certificates.

Booking checklist — send us these

  • Vessel name and IMO number
  • Instrument list from your PMS: tag, type, range, location
  • Alarm and shutdown set points for switches and transmitters
  • Berth or anchorage, ETA and ETD
  • Any instruments already known to be faulty
Send via Contact Form
Ship engineer reviewing a pressure and temperature instrument calibration certificate
Why AES

Why Ship Managers Use AES for Instrument Calibration

  • Marine-first. We calibrate instruments where they work, on board, and check the alarms they feed.
  • Quality system that covers it. ISO 9001:2015 certified, with a scope that includes ship fixed machinery instruments and control automation.
  • Safe work on board. ISO 45001:2018, bizSAFE Level 4 completed and bizSAFE Star.
  • Alongside and at anchorage. Planned around your port stay and isolation windows.
  • More done per port call. Combine with gas detector, ODME and OWS or UTI meter calibration.

See all our marine calibration services in Singapore.

Marine engineer checking temperature and pressure readings in a ship's engine control room
FAQ

Pressure & Temperature Calibration in Singapore: Common Questions

How often should ship pressure gauges and temperature sensors be calibrated?

Follow the maker's recommendation and your planned maintenance system. Many operators work to a yearly cycle, with safety-critical switches and test gauges checked more often. Our interval table lists common intervals.

Can you calibrate instruments without removing them?

Often, yes. Many gauges and transmitters can be tested in place through a test connection or isolating valve. Where an instrument has to come off, we refit it the same day.

Do you check the alarm or shutdown as well as the sensor?

Yes. For switches and transmitters we confirm the set point against your alarm list and check that the alarm or shutdown actually responds.

Are your certificates SAC-SINGLAS accredited?

Our certificates are traceable calibration certificates issued under our ISO 9001:2015 quality system. If your class, flag or charterer specifically asks for an ISO/IEC 17025 accredited certificate for an instrument, tell us when you book and we'll advise on the right route. Our article on ISO 17025 vs manufacturer calibration explains when each is accepted.

Can you calibrate at anchorage?

Yes. We attend ships at Singapore anchorages as well as alongside at terminals and shipyards. Your agent usually arranges the launch.

What happens if an instrument fails calibration?

We record the as-found reading, adjust it if the design allows, and flag it for replacement if not. The certificate shows the result either way, so your PMS stays accurate.

How much does pressure and temperature calibration cost?

It depends on the number of instruments, their types and where the ship is. Send your list through the contact form or WhatsApp and we'll price it before we board.

Instruments due this port call?

Get Your Pressure and Temperature Instruments Calibrated in Singapore

Send the vessel details and your instrument list. We'll confirm timing and price, then attend alongside or at anchorage.

Email sales@alrightes.com · 3791 Jalan Bukit Merah #06-01, E-Centre @ Redhill, Singapore 159471